Home » Areas of Care » Compounding Pharmacy » OneDose™ Single Use Ophthalmic Dropper

OneDose™ Single Use Ophthalmic Dropper

The most economical single use ophthalmic dropper available today, OneDose™ is the perfect solution for preservative-free medication dispensing while staying compliant to “beyond use dating” via USP <797>.

This style of sterile eye dropper is ideal for small batch, single dose, preservative free compounding. The OneDose™ eye droppers come in 1.5 mL with matching cap.

Cost Effective for Single Use

OneDose™ single use ophthalmic droppers are designed for compounders looking for a cost effective preservative free medication dispensing solution while remaining compliant with “beyond use dating” via USP <797>.

  • More economical than using a standard multi-use 1.5 mL bottle
  • No specialized equipment required to seal
3 layer packaging

Sterilized and Endotoxin

Each lot is released sterile and tested for endotoxin through one of our qualified third-party testing labs

All lots will be accompanied with a detailed Certificate of Analysis

 

Measured Cleanliness USP <788> Particulate Tested

Each lot is measured for particulates via a USP <788> methodology through one of our qualified third-party testing labs as follows:

All lots will be accompanied with a detailed Certificate of Analysis showing the measured cleanliness of the product.

Particulate matter test

Easy Seal Cap

  • Sterile easy seal cap comes packaged separately for easy assembly
  • Friction-fit cap passes container closure integrity testing per USP <1207> vacuum decay leak rest method
Syringes

Robust Documentation for your Quality Records

Every Lot Release is accompanied by a detailed Certificate of Analysis which includes certifications and measurements of:

  • Sterility
  • Endotoxin
  • USP <788> level particulates

If you need something above and beyond what are documentation has to offer, we can work with you to get it.

Certificate of Analysis

Product Offerings

Frequently Asked Questions

What final release testing is performed for these bottle products?
Test Method
Sterility Direct Transfer Method Embedded Spore Strip accredited to the ISO 17025
Endotoxin/LAL ISO 17025 – (USP), General Chapter <85>, (USP), General Chapter <161>, Medical Devices
USP <788> Test Method 1 – Particulates Test Method Acceptance Criteria: Based on the methodology of USP <788>

All testing is performed by qualified third party testing labs contracted with MPS Pharma. The final measurements are all reported on our current Certificate of Analysis.

Why do you measure particulates to USP <788>?

Our bottles are most used for topical ophthalmic preparations in the form of eye drops. This is a form of extraocular application of medication to the eyes. Specific chapters in USP guidelines state as follows:

USP <771> Opthalmic Products – Quality Tests:

“Two general categories apply for product administration to the tissues in the eye. Intraocular administration includes all ophthalmic products that cross (penetrate) boundary tissue, such as the cornea and sclera. For subvisible content, USP guidance is followed. Products for intraocular use must comply with Particulate Matter in Ophthalmic Solutions <789>. Product for extraocular use must comply with Particulate Matter in Injections <788>.”

Both USP <788> and <789> are applicable to finished drug products; containers are not considered finished drug product. The testing MPS Pharma has performed on the bottles is based on the methodology of USP <788>, with particulate measurements reported for 10 um and 25 um sizes against limits reported on a per container basis.

Can I see a sample Certificate of Analysis that comes with this product?

See the attached PDF for a sample Certificate of Analysis

The bottles come pre-assembled. What is the advantage of this?

Some pharmacy compounders find it advantageous to have the tip, cap, bottle come together. With special technique, the tip can be removed from the bottle just using the cap. In this manner a compounder can fill the bottle and close it with never physically touching the dropper tip.

If I open the pouch, will the bottle contents still be sterile?

Outside the bottle:

Once the pouch has been opened, then by the definition the outside of the bottle cannot be considered sterile, unless the pouch is opened in a controlled sterile environment.

Inside the bottle:

  • Inside of the bottle of an opened pouch is sterile and will remain sterile if pouches are opened in a clean and sterile environment. To maintain sterility, pouches should be handled and closed in the same controlled environment.
  • MPS does not own any specific drugs of its own, and as a result has performed the CCIT test with water. MPS bottles pass the Container Closure Integrity Testing (CCIT) as defined by USP <1207>.
  • Many clients of MPS have performed CCIT protocols against the bottle combined with their specific drugs and they have successfully passed.
  • CCIT is not a sterility test. The CCIT methodology evaluates the container’s ability to maintain a container closure system in protecting the internal contents against potential external contamination.

The passing of these tests provides additional confidence that no external contaminants can enter the closure system.

What is the drop size for these bottles?

Drop size is dependent on several factors including the product being dispensed and the technique in which it is dispensed. The best advice for customers is to test the dropper tip with the compounded product to determine its suitability in its application.

A general water reference point can be provided to our customers as a starting guideline:

Size(s) Drop Size
3 mL 42 ul +/- 5 uL
7, 10, 15, 30 mL 40 ul +/- 5 uL
What are the bottles made of? Can more information be shared?

The bottle and tip itself are composed of low-density polyethylene resin (LDPE). These are the fluid contact surfaces. The cap itself is composed of polypropylene (PP) and can be considered a non-fluid contact surface.

A more detailed material profile is available to share with customers who complete a non-disclosure agreement.

Are dimensional drawings of the bottle available?

Detailed dimensional drawings are available to customers.

Can I get a specification sheet of this item for my records?

See the attached document for the specification sheet

Our pharmacy follows USP <797> and it requires we use depyrogenated materials. Are your products depyrogenated? Do you have a Certificate of Analysis (C of A) stating such?

All MPS lots are tested for endotoxins (pyrogens) and will meet the materials requirements as defined in USP <797>. A detailed explanation of this can be found in this PDF link. All MPS C of As include measurements for these endotoxins.

Have any questions?

Contact our Customer Service team for additional support at
(844) 641-3814 or by email at [email protected]